Product description
Sensitive detection of classical rabies virus and bat-associated lyssaviruses (especially EBLV-1, EBLV-2) by direct immunofluorescence test.
Your benefits
• Immunofluorescence test based on FITC-labelled monoclonal antibodies
• Sensitive detection of classical rabies virus and bat-associated lyssavirus
• Official approval and batch testing by the National Reference Centre (FLI)
applications
Our ‘Monoclonal Anti-Rabies, FITC’ is used to detect the rabies virus in investigation material of different origins. In the direct immunofluorescence test, rabies viruses can be detected, for example, in leaching preparations of animal brains or in cell culture after virus isolation. The product contains fluoresceinisothiocyanate (FITC)-labelled monoclonal antibodies.
Application and further information
Sequence diagram
- preparation
Preparation of the use dilution of Monoclonal Anti-Rabies, FITC with H2O dest. and PBS buffer. - Fix sample
Position the sample on a slide and draw 3 times briefly through the non-luminous flame of a Bunsen burner for fixing. - Overlaying
Fixed samples with the use dilution of Monoclonal Anti-Rabies, FITC overlay. - Influence
Store slides in a damp chamber at 37 °C for 30 minutes. - Washing
Wash slides by immersion in PBS buffer 3 x 5 min each. - drying
Rinse slides with water and let them air dry. - Read the result and check for fluorescence
In the fluorescence microscope, examine the sample for fluorescence: Positive sample clearly shows tissue-differentiable, almost round particles of different sizes with intense yellow-green fluorescence and with mostly randzonal amplification of fluorescence. Tissue background must have only a weak, gray-green fluorescence.
Interpretation of the results
The preparations are examined under a microscope with a suitable fluorescence device, if necessary using the oil immersion.
Positive result
Positive samples clearly show tissue-differentiable, nearly round particles of different sizes with intense yellow-green fluorescence, often with a randzonal amplification of fluorescence. The tissue background must have only a faint grey-green fluorescence.
Negative result
The absence of particles with intense yellow-green fluorescence in a faint gray-green background indicates the absence of rabies virus.
characteristics
| contents | 1 ml |
|---|---|
| Storage temperature | 2-8 °C |
| shape | lyophilisate |
| composition | contains fluoresceinisothiocyanate-labelled monoclonal antibodies isolated from cell culture supernatants |